TOP > Various Settings of the Product > The Product's [Settings] Menu > [Printing] Menu > [Advanced] > [Calibrate] > [GAP Slice Level] [GAP Slice Level] Set the Gap sensor slice level. The setting procedure is as follows: 1. Use the following formula to calculate the slice level: (High level - Low level) x 0.3 + Low level = slice level 2. Select the [GAP Slice Level] in the [Calibrate] menu and press the button. 3. Press the / buttons to change the [Slice level] value. Set the [Slice level] to the level calculated in step 1. 4. Press the button to confirm the value. • If you set the [Slice level] to 0.0 (V), the product sets the slice level automatically. Back to top